FDA recall record

Siemens Luminos dRF Fluoroscopic X-Ray system Fluoroscopic X-Ray system recall — FDA Z-1848-2013

Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may…

Class IIclassification
Z-1848-2013official record ID
Aug 7, 2013report date

Recall record at a glance

Official recordZ-1848-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Medical Solutions USA, Inc.
Report dateAug 7, 2013
Recall initiation dateMay 7, 2013
Quantity4
LocationMalvern, PA, United States

Official product description

Siemens Luminos dRF Fluoroscopic X-Ray system Fluoroscopic X-Ray system

Why the product was recalled

Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may occur causing the table to tilt into an upright position and may collide with the floor. Four consignees were notified of this recall by Update Instruction XP067/12/S -- all four units have been corrected.

Distribution information

US Distribution including the states of CA, IN, NJ and OK.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.