SIEMENS IMMULITE 2000/2000 Xpi Systems Intact PTH (Parathyroid Hormone) Test Code iPTH recall — FDA Z-0979-2017

Siemens is conducting this recall for the IMMULITE 2000/IMMULITE 2000 XPi Intact PTH (Intact Parathyroid Hormone) (iPTH) assay kit lot 320. Siemens has…

Class IIclassification
Z-0979-2017official record ID
Jan 18, 2017report date

Recall record at a glance

Official recordZ-0979-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Healthcare Diagnostics, Inc.
Report dateJan 18, 2017
Recall initiation dateNov 22, 2016
Quantity2728 kits (US: 565 kits; OUS: 2163 kits)
LocationTarrytown, NY, United States

Official product description

SIEMENS IMMULITE 2000/2000 Xpi Systems Intact PTH (Parathyroid Hormone) Test Code iPTH, REF/Catalog Number L2KPP2, Siemens Material Number (SMN) 10381441 and REF/Catalog Number L2KPP6, SMN 10381442; IVD --- Made in UK -- Siemens healthcare Diagnostics Products Ltd. Glyn Rhonwy, Llanberis, Gwynedd, LL55 4EL UK For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers - for the quantitative measurement of intact parathyroid hormone (parathyrin, PTH) in EDTA plasma and serum, as an aid in the differential diagnosis of hypercalcemia and hypocalcemia.

Why the product was recalled

Siemens is conducting this recall for the IMMULITE 2000/IMMULITE 2000 XPi Intact PTH (Intact Parathyroid Hormone) (iPTH) assay kit lot 320. Siemens has confirmed that IMMULITE 2000/IMMULITE 2000 XPi Intact PTH kit lot 320 can exhibit an average negative bias of up to -39% at iPTH concentrations <20 pg/mL with serum and EDTA patient samples vs. a reference kit lot.

Distribution information

Worldwide Distribution: US (Nationwide) and countries of: Canada, China, Chile, Mexico, Australia, Austria, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria, Peru, Colombia, Ecuador, Bolivia, Costa Rica, Honduras, Croatia, Czech Republic, Denmark, Egypt, Estonia, France, Algeria, Greece, Guatemala, Hungary, India, Ireland, Israel, Italy, Japan, Kazakhstan, Latvia, Lebanon, Netherlands, Norway, Pakistan, Panama, Paraguay, Poland, Portugal, Republic Korea, Romania, Russian Fed., Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, Jordan, U.A.E., United Kingdom and Uruguay.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.