FDA recall record

Siemens Digital Linear Accelerators of type: ARTISTE MV System MEVATRON M2/Primus Mid-Energy recall — FDA Z-1204-2019

Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to…

Class IIclassification
Z-1204-2019official record ID
May 1, 2019report date

Recall record at a glance

Official recordZ-1204-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Medical Solutions USA, Inc.
Report dateMay 1, 2019
Recall initiation dateMar 20, 2019
Quantity120 US units
LocationMalvern, PA, United States

Official product description

Siemens Digital Linear Accelerators of type: ARTISTE MV System MEVATRON M2/Primus Mid-Energy PRIMUS PRIMUS HI ONCOR Impression ONCOR Impression Plus ONCOR Avant Garde ONCOR Expression ARTISTE, ONCOR, and PRIMUS running the following: Control Console from software version 13.0.302 and higher; Control Console from software version 9.2.400 and higher; and Control Console from software version 11.0.400 and higher.

Why the product was recalled

Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field Sequence (AFS) Automatic Motion Protection (AMP) function implemented at the control of the Digital Linear Accelerator.

Distribution information

Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.