Official product description
Siemens Digital Linear Accelerators of type: ARTISTE MV System MEVATRON M2/Primus Mid-Energy PRIMUS PRIMUS HI ONCOR Impression ONCOR Impression Plus ONCOR Avant Garde ONCOR Expression ARTISTE, ONCOR, and PRIMUS running the following: Control Console from software version 13.0.302 and higher; Control Console from software version 9.2.400 and higher; and Control Console from software version 11.0.400 and higher.
Why the product was recalled
Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field Sequence (AFS) Automatic Motion Protection (AMP) function implemented at the control of the Digital Linear Accelerator.
Distribution information
Nationwide
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.