Recall record at a glance
| Official record | Z-1039-2019 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Shimadzu Medical Systems Usa Com |
| Report date | Mar 27, 2019 |
| Recall initiation date | Jan 14, 2019 |
| Quantity | 93 units |
| Location | Torrance, CA, United States |
Official product description
SHIMADZU MobileDaRt Evolution (MX8 Version), Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or general radiography of patients in the hospital who cannot move and/or of outpatients in emergency. Exposure can be captured on film, CR or Digital Radiography (DR).
Why the product was recalled
In these units, the brake is usually released by gripping the drive handle and applied by letting go of the handle. However, for systems within a specified manufacture period, we have identified that even when letting go of the handle, the systems brakes are not applied. In this event, if the Operator releases the handle when the device is moving, the unit will continue to coast.
Distribution information
US Nationwide Distribution
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.