FDA recall record

Shape Medical Systems Shape-HF Cardiopulmonary Testing System Disposable Patient Interface recall — FDA Z-1060-2015

This recall has been initiated because use of the impacted DPIs could result in erroneous testing results potentially leading to incorrect diagnosis and…

Class IIclassification
Z-1060-2015official record ID
Feb 11, 2015report date

Recall record at a glance

Official recordZ-1060-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmShape Medical Systems, Inc.
Report dateFeb 11, 2015
Recall initiation dateDec 23, 2014
Quantity564 devices
LocationSaint Paul, MN, United States

Official product description

Shape-HF Cardiopulmonary Testing System Disposable Patient Interface (DPI). Part Number 0004-4001. The Shape-HF Cardiopulmonary Exercise Testing System is a stationary device that monitors parameters during laboratory or clinical conditions.

Why the product was recalled

This recall has been initiated because use of the impacted DPIs could result in erroneous testing results potentially leading to incorrect diagnosis and incorrect treatment. The magnitude of error for VO2 or VCO2 parameters could be approximately 20% with use of the impacted product.

Distribution information

Distributed in the states of AZ, FL, GA, MA, MN, NJ. NY, PA, TX, and WI.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.