FDA recall record

CareFusion 213 Sepp, Compound Benzoin Tincture Packaged and Tincture 3,000 Bulk recall — FDA Z-1001-2013

An internal review of CareFusion's labeling for preoperative skin prep product codes was recently conducted. The labeling of the product codes identified…

Class IIIclassification
Z-1001-2013official record ID
Apr 3, 2013report date

Recall record at a glance

Official recordZ-1001-2013
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmCareFusion 213, LLC
Report dateApr 3, 2013
Recall initiation dateMar 4, 2013
Quantity13,749,000
LocationEl Paso, TX, United States

Official product description

Sepp, Compound Benzoin Tincture Packaged and Tincture 3,000 Bulk. Product Usage: Preoperative skin product

Why the product was recalled

An internal review of CareFusion's labeling for preoperative skin prep product codes was recently conducted. The labeling of the product codes identified above state that the contents of the package are sterile and does not clarify that this sterility claim is only applicable to the applicator. The antiseptic solution that is applied to the skin is a cutaneous disinfectant, and is not sterilized.

Distribution information

USA Nationwide Distribution

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

DrugJul 8, 2026

ChloraPrep FREPP, Clear recall — FDA D-0646-2026

Lack of assurance of sterility: Unsterilized ChloraPrep & PurPrep Applicators, intended for further processing and sterilization were instead distributed to customers outside of the…

Class IIFDA: OngoingCareFusion 213, LLC