FDA recall record

Nonin Medical SenSmart, Regional Oximetry EQUANOX Classic Plus Adult/Pediatric Sensor > 40 recall — FDA Z-0901-2015

Nonin Medical is conducting a recall of certain regional oximetry sensors and associated products. They have received reports of signal loss during use of…

Class IIclassification
Z-0901-2015official record ID
Jan 7, 2015report date

Recall record at a glance

Official recordZ-0901-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNonin Medical, Inc.
Report dateJan 7, 2015
Recall initiation dateNov 25, 2014
Quantity5892 (5452 US, 440 OUS)
LocationPlymouth, MN, United States

Official product description

SenSmart, Regional Oximetry EQUANOX Classic Plus Adult/Pediatric Sensor > 40 kg., Model 8003CA (REF 9824-001, 3-Wavelength), 8004CA (REF 9705-001, 4-Wavelength), with associated materials (INT-100 cable, operator's manual; and Quick Start Guide). Sensors are single-packed in a pouch, then multi-packed, 20 per shelf-box.

Why the product was recalled

Nonin Medical is conducting a recall of certain regional oximetry sensors and associated products. They have received reports of signal loss during use of our shortcabled regional oximetry sensors with diaphoretic patients. No injury or illness have been reported.

Distribution information

To be provided later.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.