FDA recall record

Siemens Medical Solutions USA Sensis/ Sensis Vibe systems with software VD1X as follows recall — FDA Z-0226-2022

System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the…

Class IIclassification
Z-0226-2022official record ID
Nov 24, 2021report date

Recall record at a glance

Official recordZ-0226-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSiemens Medical Solutions USA, Inc.
Report dateNov 24, 2021
Recall initiation dateOct 5, 2021
Quantity403 units
LocationMalvern, PA, United States

Official product description

Sensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo Model 11007641 (3)Sensis Vibe Combo Model 11007642 Intended Use: A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies

Why the product was recalled

System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances, this may result in a partial freeze of the user interface which would no longer update the vital signs

Distribution information

Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.