Recall record at a glance
| Official record | Z-1434-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Siemens Medical Solutions USA, Inc. |
| Report date | Apr 28, 2021 |
| Recall initiation date | Mar 4, 2021 |
| Quantity | 99 units |
| Location | Malvern, PA, United States |
Official product description
Sensis/ Sensis Vibe Hemo systems with VD12A software as follows: Dedicated SIS Server- 6648153 Sensis Post-Processing Workstation- 6648161 Sensis- 10764561 VM Virtual Server- 10765502 Sensis High-End Servers- 10910620 Sensis Vibe Hemo- 11007641 Sensis Vibe Combo- 11007642 Sensis Vibe is a recording and procedure data management system for interventional cardiology, interventional radiology, surgical procedures performed in a hybrid OR and electrophysiology
Why the product was recalled
Due to the configuration of certain Windows Service Permissions within the operating systems of the Sensis/ Sensis Vibe computer, there is a risk for exposure of sensitive information, manipulation of data, or Denial of Service attacks and could result in incorrect diagnostic or therapeutic decisions
Distribution information
US Nationwide distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.