FDA recall record

Datascope SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to recall — FDA Z-2221-2023

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks…

Class IIclassification
Z-2221-2023official record ID
Aug 2, 2023report date

Recall record at a glance

Official recordZ-2221-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmDatascope Corp.
Report dateAug 2, 2023
Recall initiation dateJun 23, 2023
Quantity321,609 total kits
LocationFairfield, NJ, United States

Official product description

SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES D684-00-0568-01 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (W/O STATLOCK) D684-00-0568-03 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (WITH PRESSURE D684-00-0568-05 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (TURKEY) D684-00-0568-09 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES, US ONLY D684-00-0568-01U PACKAGED INSERTION KIT - SENSATION PLUS 7.5 Fr. 40 cc IAB D884-00-0019-22

Why the product was recalled

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Distribution information

Domestic distribution nationwide. International distribution worldwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: N/A.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMar 25, 2026

Datascope Battery Charging Station recall — FDA Z-1587-2026

The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has…

Class IIIFDA: OngoingDatascope Corp.