FDA recall record

Hologic Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version recall — FDA Z-3248-2024

Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the…

Class IIclassification
Z-3248-2024official record ID
Oct 2, 2024report date

Recall record at a glance

Official recordZ-3248-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmHologic Inc.
Report dateOct 2, 2024
Recall initiation dateAug 19, 2024
Quantity425 units
LocationNewark, DE, United States

Official product description

Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version Dimensions 1.12.0 and 3Dimensions 2.3.0 system software as follows Model/Catalog Number: 1. ASY-16824 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.1 SYSTEMS 2. ASY-16825 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.2.1 SYSTEMS 3. ASY-17219 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.10.0 SYSTEMS 4. ASY-17220 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.1.0 SYSTEMS 5. ASY-17221 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.0 SYSTEMS 6. ASY-17222 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.2.0 SYSTEMS 7. 3DM-LIC-SW-UP-V2.3 DIMENSIONS SOFTWARE 2.3 (3DIMENSIONS) 8. SDM-LIC-SW-UP-V1.12 DIMENSION SOFTWARE 1.12 (SELENIA DIMENSIONS)

Why the product was recalled

Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis

Distribution information

Worldwide distribution - US Nationwide and the countries of Belgium, Canada, Israel, Korea, Republic of, Netherlands.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceFeb 19, 2025

Sertera Biopsy Device recall — FDA Z-1170-2025

Devices presented a condition in which the inner needle separated from the hub and produced a longer throw of the inner needle or dislodgement.

Class IIFDA: OngoingHologic, Inc.