Recall record at a glance
| Official record | Z-3143-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | SEER MEDICAL PTY LTD |
| Report date | Sep 25, 2024 |
| Recall initiation date | Aug 12, 2024 |
| Quantity | 3,948 units |
| Location | Melbourne, Australia |
Official product description
Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and video EEG (vEEG) recordings and analysis. The Seer Home System includes the Sight Monitoring Hub, the Sense wearable, Monitoring Portal and associated accessory components (Disposable EEG Wires, WaterTabs, WaterTabs Application Kit, and Take-home Kit).
Why the product was recalled
During CAPA investigations of sight production inconsistencies, it was highlighted that previous Electromagnetic Compatibility (EMC) assessments related to RF radiated emissions may not be reasonably equivalent to the current product due to design changes in the system. A new CAPA was opened to investigate the radiated emissions compliance issues. - Seer Home Systems distributed and retained in the field were not compliant with Electromagnetic Compatibility (EMC) testing according to IEC 60601-1-2 - Seer Home Systems, including accessories, were manufactured and distributed without application of proper design change and manufacturing controls - Seer Home Systems were distributed with labeling deficiencies
Distribution information
Worldwide - US Nationwide distribution in the states of MN, NY, PA and the countries of Australia.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.