FDA recall record

POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC SEDIPLAST; Model/Catalog Numbers: S-100 and recall — FDA Z-0107-2027

During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the…

Class IIclassification
Z-0107-2027official record ID
Oct 14, 2026report date

Recall record at a glance

Official recordZ-0107-2027
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPOLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC
Report dateOct 14, 2026
Recall initiation dateAug 25, 2026
Quantity448400
LocationCortlandt Manor, NY, United States

Official product description

SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;

Why the product was recalled

During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).

Distribution information

Worldwide - US Nationwide distribution in the states and territory of AK, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and the country of United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.