FDA recall record

Ethicon Endo-Surgery SEDASYS Computer-Assisted Personalized Sedation System, SEDPRU01 recall — FDA Z-2374-2015

Ethicon has found that disinfecting methods not specified in the Instructions For Use (IFU) have led to the degradation of the plastic material of the…

Class Iclassification
Z-2374-2015official record ID
Sep 2, 2015report date

Recall record at a glance

Official recordZ-2374-2015
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmEthicon Endo-Surgery Inc.
Report dateSep 2, 2015
Recall initiation dateMay 13, 2015
Quantity14 units
LocationBlue Ash, OH, United States

Official product description

SEDASYS Computer-Assisted Personalized Sedation System, SEDPRU01. An integrated physiological monitoring and drug delivery system intended to provide physician-led teams with a means of administering 1% propofol (10 mg/mL) injectable emulsion to sedate patients > 18 (ASA I-II) undergoing esophagogastroduodenoscopy (EGD) and colonoscopy procedures. Intended to provide physician-led teams with a means of administering 1% propofol (10 mg/mL) injectable emulsion to sedate patients > 18 (ASA I-II) undergoing esophagogastroduodenoscopy (EGD) and colonoscopy procedures.

Why the product was recalled

Ethicon has found that disinfecting methods not specified in the Instructions For Use (IFU) have led to the degradation of the plastic material of the SEDASYS System Control Unit.

Distribution information

Nationwide Distribution including OH, RI, TX & WA.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.