FDA recall record

Micromed S.p.A SD LTM 64 PLUS, EEG Amplifier/recorder recall — FDA Z-2353-2026

Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a…

Class IIclassification
Z-2353-2026official record ID
Jun 17, 2026report date

Recall record at a glance

Official recordZ-2353-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMicromed S.p.A.
Report dateJun 17, 2026
Recall initiation dateMay 4, 2026
Quantity207 units
LocationMogliano Veneto, N/A, Italy

Official product description

SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02

Why the product was recalled

Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.

Distribution information

Worldwide - US Nationwide distribution in the states of FL, PA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.