FDA recall record

Zimmer Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally recall — FDA Z-0285-2025

A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail…

Class IIclassification
Z-0285-2025official record ID
Nov 13, 2024report date

Recall record at a glance

Official recordZ-0285-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmZimmer, Inc.
Report dateNov 13, 2024
Recall initiation dateSep 17, 2024
Quantity432 units
LocationWarsaw, IN, United States

Official product description

Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. Item Number: VPX4000 (7273)

Why the product was recalled

A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery

Distribution information

Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0972-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0970-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0975-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.
Medical DeviceDec 31, 2025

Affixus Antegrade Femoral Nails recall — FDA Z-0968-2026

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs…

Class IIFDA: OngoingZimmer, Inc.