Recall record at a glance
| Official record | Z-0208-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Cardinal Health 200, LLC |
| Report date | Oct 29, 2025 |
| Recall initiation date | Sep 18, 2025 |
| Quantity | all lots |
| Location | Waukegan, IL, United States |
Official product description
Salem Sump PVC Tubes: Product Code/Product Name: 8888265108 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 10 Fr/Ch (3.3 mm), 36 (91 cm) 8888265116 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 12 Fr/Ch (4.0 mm), 48 (122 cm) 8888265124 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 14 Fr/Ch (4.7 mm), 48 (122 cm) 8888265132 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 16 Fr/Ch (5.3 mm), 48 (122 cm) 8888265140 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 18 Fr/Ch (6.0 mm), 48 (122 cm) 8888266213 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 10 Fr/Ch (3.3 mm) x 36 (91 cm) 8888266221 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 12 Fr/Ch (4.0 mm) x 48 (122 cm) 8888266247 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 14 Fr/Ch (4.7 mm) x 48 (122 cm) 8888266262 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 16 Fr/Ch (5.3 mm) x 48 (122 cm) 8888266288 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 18 Fr/Ch (6.0 mm) x 48 (122 cm)
Why the product was recalled
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
Distribution information
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.