FDA recall record

Cardinal Health 200 Salem Sump PVC Tubes: Product Code/Product Name: 8888264911E ENFit" recall — FDA Z-0206-2026

Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use.…

Class IIclassification
Z-0206-2026official record ID
Oct 29, 2025report date

Recall record at a glance

Official recordZ-0206-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCardinal Health 200, LLC
Report dateOct 29, 2025
Recall initiation dateSep 18, 2025
Quantityall lots
LocationWaukegan, IL, United States

Official product description

Salem Sump PVC Tubes: Product Code/Product Name: 8888264911E ENFit" SALEM SUMP" PVC TUBE 10FR 36IN 8888264929E ENFit" SALEM SUMP" PVC TUBE 12FR 48IN 8888264945E ENFit" SALEM SUMP" PVC TUBE 14FR 48IN 8888264960E ENFit" Salem Sump" PVC Tube 16Fr 48IN 8888264986E ENFit" SALEM SUMP" PVC TUBE 18FR 48IN 8888266106E PVC SALEM SUMP WITH ENFIT & ARV 10FR 36IN 8888266114E PVC SALEM SUMP WITH ENFIT & ARV 12FR 48IN 8888266122E PVC SALEM SUMP WITH ENFIT & ARV 14FR 48IN 8888266130E PVC SALEM SUMP WITH ENFIT & ARV 16FR 48IN 8888266148E PVC SALEM SUMP WITH ENFIT & ARV 18FR 48IN 8888268060E ENFit" SALEM SUMP" PVC TUBE 6FR 24IN 8888268086E ENFit" Salem Sump" PVC Tube 8Fr 24IN

Why the product was recalled

Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

Distribution information

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.