FDA recall record

Medline Industries Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry recall — FDA Z-1320-2026

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory…

Class Iclassification
Z-1320-2026official record ID
Feb 18, 2026report date

Recall record at a glance

Official recordZ-1320-2026
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmMedline Industries, LP
Report dateFeb 18, 2026
Recall initiation dateDec 22, 2025
Quantity311
LocationNorthfield, IL, United States

Official product description

Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 5mm 81107RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 2-5-2mm 81402RH 81404RH 81405RH 81474RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81473RH

Why the product was recalled

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

Distribution information

Distribution US nationwide.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.