FDA recall record

Nobel Biocare Usa Replace Select Tapered TiU NP 3.5x13mm, Article No. 29402 recall — FDA Z-1415-2015

A material used in the packaging of the dental implants may impart a sticky substance on the top surfaces of the implants and in some cases on the first…

Class IIclassification
Z-1415-2015official record ID
Apr 22, 2015report date

Recall record at a glance

Official recordZ-1415-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNobel Biocare Usa LLC
Report dateApr 22, 2015
Recall initiation dateMar 9, 2015
Quantity17 units
LocationYorba Linda, CA, United States

Official product description

Replace Select Tapered TiU NP 3.5x13mm, Article No. 29402. Nobel Biocare's Dental Implants are root-form endosseous implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's Dental Implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. Dental Implants may be placed immediately and put into immediate function providing that the initial stability requirements detailed in the surgical manuals are satisfied.

Why the product was recalled

A material used in the packaging of the dental implants may impart a sticky substance on the top surfaces of the implants and in some cases on the first threads.

Distribution information

Worldwide Distribution - US (nationwide) and Internationally to Austria, Australia, Belgium, Bulgaria, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Lithuania, Luxembourg, Mexico, Netherlands, Norway, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Sweden, Turkey, and United Arab Emirates.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.