Cordis RenLane Renal Denervation Catheter recall — FDA Z-2405-2015

Potential Damage may occur to the helical tip of the RENLANE Renal Denervation Catheter, if the hemostasis valve of the arterial access device is not…

Class IIclassification
Z-2405-2015official record ID
Aug 26, 2015report date

Recall record at a glance

Official recordZ-2405-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCordis Corporation
Report dateAug 26, 2015
Recall initiation dateApr 15, 2014
Quantity98 units
LocationMiami Lakes, FL, United States

Official product description

RenLane Renal Denervation Catheter; Cat No.: D135601. Used in adults patients (>18 years) with drug resistant hypertension to denervate the renal arteries to reduce blood pressure.

Why the product was recalled

Potential Damage may occur to the helical tip of the RENLANE Renal Denervation Catheter, if the hemostasis valve of the arterial access device is not fully open before insertion or withdrawal of the catheter.

Distribution information

Distributed in Germany only.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.