Recall record at a glance
| Official record | Z-0582-2016 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Vidco, Inc. |
| Report date | Jan 13, 2016 |
| Recall initiation date | Nov 8, 2015 |
| Quantity | 59 units |
| Location | Beaverton, OR, United States |
Official product description
Remote Patient Monitoring System, MDP 2000 Series, Part Numbers: MDP2040-0100, PGM340R3, or PGM340R4. The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected Examples of areas where the MDP200 is used include ICU, CCU, PACU, Emergency, Telemetry Step-down and other areas where patient's physiological information are to be observed at one or more locations. Typical information displayed includes ECG and blood pressure waveforms and numeric values which as heart rate and systolic, mean and diastolic pressure. This device acts as the Central Monitor Station, it permits from one (1) to sixteen (16) bedside monitor(s) to be connected simultaneously.
Why the product was recalled
Testing at customer site showed unit Remote Patient Monitoring System MDP2040-0100 in a continuous trap condition, not allowing system to reset and reboot. Two customers recently complained of the system freezing and it could only be restarted if the user re-applied power.
Distribution information
US Nationwide distribution in the states of AZ, CA, MD, NM, NJ, and OH.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.