Steris Reliance 1227 Cart and Utensil Washer/Disinfector, FW03101, FW03102, FW03S003 Product recall — FDA Z-2392-2018

The firm has become aware that the Reliance 1227 Cart and Utensil Washer/Disinfector s Chemical Low Level alarm, intended to prevent the user from…

Class IIclassification
Z-2392-2018official record ID
Jul 18, 2018report date

Recall record at a glance

Official recordZ-2392-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSteris Corporation
Report dateJul 18, 2018
Recall initiation dateMay 23, 2018
Quantity118
LocationMentor, OH, United States

Official product description

Reliance 1227 Cart and Utensil Washer/Disinfector, FW03101, FW03102, FW03S003 Product Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theatre shoes and other miscellaneous reusable items used in the care of patients.

Why the product was recalled

The firm has become aware that the Reliance 1227 Cart and Utensil Washer/Disinfector s Chemical Low Level alarm, intended to prevent the user from initiating a cycle when a low chemical level situation occurs, does not operate as intended. Currently, if a low chemical level situation occurs, the alarm will only be generated at the unit s next power-up or when accessing service mode. The failure of the Reliance 1227 to identify that the chemicals used for cleaning are low or empty could result in bedpans and urinals, basins, case carts, beds, theater shoes and other miscellaneous reusable patient care items not being properly cleaned before disinfection or reuse.

Distribution information

Worldwide Distribution - US Nationwide in the states of AL, AK, CA, CO, CN, DE, FL, GA, IL, IA, KS, KY, LA, MA, MI, NJ, NY, ND, OH, OK, OR, PA, RI, SC, TN, TX, VT, VA, WI, and WV. Foreign distribution to Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceAug 14, 2024

Steris Hookup recall — FDA Z-2538-2024

IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an…

Class IIFDA: OngoingSteris Corporation