FDA recall record

RAYSEARCH LABORATORIES AB RayStation Therapy Treatment Planning System Stand-alone Software recall — FDA Z-1712-2016

A software issue with editing tools that use the left mouse button held down, for drawing in and interacting with the patient views and beams eye views.…

Class IIclassification
Z-1712-2016official record ID
Jun 1, 2016report date

Recall record at a glance

Official recordZ-1712-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRAYSEARCH LABORATORIES AB
Report dateJun 1, 2016
Recall initiation dateFeb 10, 2016
Quantity1,264 Total units (552 units domestically & 711 units internationally)
LocationStockholm, N/A, Sweden

Official product description

RayStation Therapy Treatment Planning System Stand-alone Software 3.0, 3.5, 4.0, 4.5 and 4.7., designed for treatment planning and analysis of radiation therapy.

Why the product was recalled

A software issue with editing tools that use the left mouse button held down, for drawing in and interacting with the patient views and beams eye views. The views can become unsynchronized with the stored data if simultaneously right clicking, pressing Ctrl-S or Ctrl-Z while the left mouse button is held down. This bug does not affect dose computations, which are based on the stored system state.

Distribution information

**Domestic** AR, AZ, CA, CT, CO, DE, FL, GA, IL, LA, MI, MD, MS, MA, NY, NV, NJ, OH, OR, PA, NC, RI, TN, WA & Washington D.C; **Internationally** Algeria, Australia, Austria, Belgium, Canada, China, France, French Polynesia, Germany, Iran, Italy, Japan, Netherlands, New Zealand, Norway, Republic of Korea, Spain, Sweden, Switzerland, Thailand & United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1110-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1106-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1108-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1109-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB