FDA recall record

RAYSEARCH LABORATORIES AB RayStation stand-alone software treatment planning system recall — FDA Z-0866-2020

Three issues found: i) The Map ROI options in the ROI list in the Structure Definition module may generate unintended ROI geometries ii) Elekta guard leaf…

Class IIclassification
Z-0866-2020official record ID
Feb 5, 2020report date

Recall record at a glance

Official recordZ-0866-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRAYSEARCH LABORATORIES AB
Report dateFeb 5, 2020
Recall initiation dateDec 19, 2019
Quantity3114 Devices/Licenses
LocationStockholm, N/A, Sweden

Official product description

RayStation stand-alone software treatment planning system. Models 4.0, 4.3, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0.

Why the product was recalled

Three issues found: i) The Map ROI options in the ROI list in the Structure Definition module may generate unintended ROI geometries ii) Elekta guard leaf behavior. There is an interoperability issue with Elekta regarding setting of guard leaves. iii)SSD, when intended as source-to-surface distance, it sometimes gives source-to-skin distance.To the best of our knowledge, these issues have not caused any patient mistreatment or other incidents. However, the user must be aware of the following information to avoid incorrect dose calculations during treatment planning.

Distribution information

US Nationwide distribution in states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, VT, WA, WI

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1110-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1106-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1108-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1109-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB