FDA recall record

RAYSEARCH LABORATORIES AB RayStation Product Usage: is a software system designed for recall — FDA Z-0380-2018

For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the…

Class IIclassification
Z-0380-2018official record ID
Jan 24, 2018report date

Recall record at a glance

Official recordZ-0380-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRAYSEARCH LABORATORIES AB
Report dateJan 24, 2018
Recall initiation dateJul 17, 2017
Quantity117 units
LocationStockholm, N/A, Sweden

Official product description

RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

Why the product was recalled

For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.

Distribution information

US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1110-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1106-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1108-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1109-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB