FDA recall record

RAYSEARCH LABORATORIES AB RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, RayPlan 1 recall — FDA Z-0589-2021

Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal…

Class IIclassification
Z-0589-2021official record ID
Dec 23, 2020report date

Recall record at a glance

Official recordZ-0589-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRAYSEARCH LABORATORIES AB
Report dateDec 23, 2020
Recall initiation dateNov 11, 2020
Quantity124
LocationStockholm, N/A, Sweden

Official product description

RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, including all service packs. RayStation 10A and RayPlan 10A are also affected, but not the 10A service pack. (Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0)

Why the product was recalled

Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan 9B and 10A, 6D couch angles may be incorrect.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1110-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1106-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1108-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1109-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB