FDA recall record

RAYSEARCH LABORATORIES AB RayStation 4.0, 4.5, 4.7 and 5.0 recall — FDA Z-0721-2017

Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans…

Class IIclassification
Z-0721-2017official record ID
Dec 14, 2016report date

Recall record at a glance

Official recordZ-0721-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRAYSEARCH LABORATORIES AB
Report dateDec 14, 2016
Recall initiation dateNov 9, 2016
QuantityUS: 66 units
LocationStockholm, N/A, Sweden

Official product description

RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Why the product was recalled

Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that enter the patient surface at an oblique angle, the dose calculation accuracy may be less than expected.

Distribution information

US Nationwide Distribution in the states of California, Florida, Illinois, Louisiana, Michigan, New Jersey, Tennessee, Texas, Washington

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1110-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1106-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1108-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB
Medical DeviceJan 28, 2026

RayStation recall — FDA Z-1109-2026

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and…

Class IIFDA: OngoingRAYSEARCH LABORATORIES AB