FDA recall record

Siemens Healthcare Diagnostics RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) recall — FDA Z-0607-2024

RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could…

Class IIclassification
Z-0607-2024official record ID
Dec 27, 2023report date

Recall record at a glance

Official recordZ-0607-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSiemens Healthcare Diagnostics Inc.
Report dateDec 27, 2023
Recall initiation dateOct 26, 2023
Quantity6035 cartridges US
LocationNorwood, MA, United States

Official product description

RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 250-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491447

Why the product was recalled

RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.

Distribution information

Nationwide with Worldwide Distribution to: Albania Algeria Angola Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Brunei Darussalam Bulgaria Burkina Faso Cambodia Canada Chile China Colombia Congo, Republic Costa Rica C¿te dIvoire Croatia Cura¿ao Czech Republic Denmark Ecuador Egypt Estonia Fiji Finland France French Polynesia Georgia Germany Ghana Greece Guadeloupe Honduras Hong Kong Hungary Iceland India Indonesia Iran, Islamic Republic of Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kenya Kosovo Kuwait Latvia Lebanon Lesotho Lithuania Luxembourg Macedonia, former Yugoslav Rep. of Malaysia Maldives Mali Martinique Mexico Moldova, Rep. of Montenegro Namibia Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestinian Territory, Occupied Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Great Britain United Republic of Tanzania Uruguay Uzbekistan Vietnam

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.