Recall record at a glance
| Official record | Z-0123-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Randox Laboratories Ltd. |
| Report date | Oct 26, 2016 |
| Recall initiation date | Aug 2, 2016 |
| Quantity | 18 |
| Location | Crumlin (North), N/A, Ireland |
Official product description
Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the enzyme lipase in serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.
Why the product was recalled
According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place directly before or after LDL or Triglycerides. A correction was made to the IFU for the Lipase assay and all customers who purchased the Lipase kit in the USA were contacted with the updated instruction.
Distribution information
US Nationwide Distribution
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.