FDA recall record

Response Biomedical Randox Laboratories Ltd recall — FDA Z-1461-2020

The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls…

Class IIclassification
Z-1461-2020official record ID
Mar 18, 2020report date

Recall record at a glance

Official recordZ-1461-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmResponse Biomedical Corp.
Report dateMar 18, 2020
Recall initiation dateMar 15, 2019
Quantity543 units
LocationVancouver, Canada

Official product description

Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318) IFU: "RAMP Cardiac Controls *** CAT. No.: C2003-2 *** Size: 3 x 3 ML". Product Usage: use in the quality control of cardiac marker assays run on the RAMP platform.

Why the product was recalled

The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house monitoring, determined that liquid Cardiac Control has seen a decrease in CK-MB Levels of up to 57% signal loss within 11 months. Shelf-life of product is 24 months.

Distribution information

Worldwide distribution: US Nationwide distributions in the states of IN, GA, OK, KS, AZ, SC, NE, NC, MI, FL, CA, MN, and AR. Countries of Canada, Italy, Malaysia, Kuwait, Saudi Arabia, UAE, Ukraine, Russia, Dominican Republic, France, Mexico, Colombia, Singapore, China, Serbia, Romania, and Philippine.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.