FDA recall record

Cook Radial Artery Pressure Monitoring Catheter Sets and Trays The subject sets and trays recall — FDA Z-0802-2016

A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical…

Class Iclassification
Z-0802-2016official record ID
Mar 2, 2016report date

Recall record at a glance

Official recordZ-0802-2016
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmCook Inc.
Report dateMar 2, 2016
Recall initiation dateJan 6, 2016
Quantity12691
LocationBloomington, IN, United States

Official product description

Radial Artery Pressure Monitoring Catheter Sets and Trays The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.

Why the product was recalled

A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.

Distribution information

Worldwide Distribution - US Nationwide including DC and PR including the countries of Australia, Germany, China, Chile, Dominican Republic, Panama, Israel, Brazil, Canada

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.