FDA recall record

Sunrise Medical (US) Quickie Q7 wheelchair recall — FDA Z-2290-2016

There is an observed failure of the backrest bracket which can fatigue over the life time of the product. When this occurs there is the possibility of the…

Class IIclassification
Z-2290-2016official record ID
Aug 3, 2016report date

Recall record at a glance

Official recordZ-2290-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSunrise Medical (US) LLC
Report dateAug 3, 2016
Recall initiation dateJul 1, 2016
Quantity12,142 devices in US, 1,710 devices internationally.
LocationFresno, CA, United States

Official product description

Quickie Q7 wheelchair. Model EIR4 Physical Medicine. The Quickie Q7 series of wheelchair's intended use is to provide mobility to persons limited to a sitting position.

Why the product was recalled

There is an observed failure of the backrest bracket which can fatigue over the life time of the product. When this occurs there is the possibility of the bracket fracturing causing the backrest to detach from the chair frame.

Distribution information

US and Mexico, Japan, Chile, Kuwait, Columbia, Hong Kong, China, Uruguay, South Africa, Costa Rica, South Korea, Singapore, New Zealand, El Salvador, Bahrain, Uruguay, Lebanon, United Arab Emirates, Brazil, Argentina, Netherlands Antilles, British Virgin Islands,

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.