Recall record at a glance
| Official record | Z-2215-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Qiagen GmbH |
| Report date | Aug 2, 2023 |
| Recall initiation date | May 9, 2023 |
| Quantity | 368 kits |
| Location | Hilden, N/A, Germany |
Official product description
QIAstat-Dx Respiratory SARS-CoV-2 Panel -IVD intended for the detection and differentiation of nucleic acid from SARS-CoV-2 and the following organism types and subtypes: Adenovirus, Coronavirus 229E, Coronavirus HKU1, Coronavirus NL63, Coronavirus OC43, SARS-CoV-2, Human Metapneumovirus A+B, Influenza A, Influenza A H1, Influenza A H3, Influenza A H1N1/pdm09, Influenza B, Parainfluenza virus 1, Parainfluenza virus 2, Parainfluenza virus 3, Parainfluenza virus 4, Rhinovirus/Enterovirus, Respiratory Syncytial Virus A+B, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae REF 691223
Why the product was recalled
If cartridges of the affected Serial numbers (SNs) are used, in most cases, the error codes 0xY0003EA and 0xY00045A (where Y is a number between 1 and 4) that lead to a run abortion will occur and no results will be reported
Distribution information
AL, AR, CA, DC, FL, GA, ID, KS, KY, MD, MN, NJ, NY,PA, SC, TX, VA
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.