FDA recall record

Philips Ultrasound Q-Station Quantification Software recall — FDA Z-0818-2015

When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may…

Class IIclassification
Z-0818-2015official record ID
Dec 24, 2014report date

Recall record at a glance

Official recordZ-0818-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPhilips Ultrasound, Inc.
Report dateDec 24, 2014
Recall initiation dateNov 5, 2014
Quantity43 units (13 in US and 30 outside US)
LocationBothell, WA, United States

Official product description

Q-Station Quantification Software. Q-Station 3.0 with a2DQ and/or aCMQ Q-Apps are affected. Model: Q-Station, Catalog number: 795088, and Part Number: 989605382391. Intended to manage, view, analyze, and report qualitative and quantitative image data from ultrasound exams.

Why the product was recalled

When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.

Distribution information

Worldwide distribution (for FCO 83000190): US, including the states of AK, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, VA, WA, and WI; and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Germany, Ecuador, Estonia, Spain, Finland, France, Great Britain, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, Norway, New Zealand, Philippines, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Turkey, Taiwan, and Vietnam. ********************************************************************************** Worldwide Distribution -- US, including the states of KY, TN, ME, CA, PA, NC, CT, NY, SC, IL, IN, and WI; and the countries of Argentina, Spain, India, Italy, Austria, Germany, Belgium, Canada, Finland, Switzerland, Greece, Malaysia, Sweden, Australia, Netherlands, Mexico, Peru, Taiwan, Japan, France, and Norway.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.