FDA recall record

Arjohuntleigh Magog Q-Connect 272 KG Scale used with Maxi Sky 2 ceiling lift The Scale is recall — FDA Z-2661-2018

The spreader bar can detach from the scale attachment under certain conditions.

Class IIclassification
Z-2661-2018official record ID
Aug 15, 2018report date

Recall record at a glance

Official recordZ-2661-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmArjohuntleigh Magog
Report dateAug 15, 2018
Recall initiation dateFeb 12, 2018
Quantity874 Units (both styles)
LocationMagog, Quebec, Canada

Official product description

Q-Connect 272 KG Scale used with Maxi Sky 2 ceiling lift The Scale is intended to be used with patient lifters to measure the weight of a patient who cannot stand on a scale. It is designed to be used indoors in hospitals, nursing homes, home care or health care facilities.

Why the product was recalled

The spreader bar can detach from the scale attachment under certain conditions.

Distribution information

Worldwide distribution: US (nationwide) to states of: CO, FL, IA, IL, KY, MN, NY, PH, and WA; and countries of: Australia, Belgium, Brazil, Canada, China, France, Germany, Hong Kong, Ireland, Italy, Japan, Netherlands, Norway, Singapore, Swiss, Turkey, United Kingdom, and United Arab Emirates.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.