Precision Valve & Automation PVA, PREVENT is an FDA-authorized emergency ventilator designed recall — FDA Z-2805-2020

While operating the machine in "Run" mode an unexcepted event may occur during the error-checking for stepper movements and pressure values. The result of…

Class IIclassification
Z-2805-2020official record ID
Aug 26, 2020report date

Recall record at a glance

Official recordZ-2805-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPrecision Valve & Automation, Inc.
Report dateAug 26, 2020
Recall initiation dateJun 26, 2020
Quantity3 units
LocationCohoes, NY, United States

Official product description

PVA, PREVENT is an FDA-authorized emergency ventilator designed to treat patients suffering from COVID-19 respiratory failure. This motorized device automatically compresses a resuscitator bag to provide air to the lungs of a COVID-19 patient, eliminating the human error and fatigue that results from manual actuation. PREVENT is indicated for emergency resuscitation with appropriate patient monitoring on adult patients that require mechanical respiratory support.

Why the product was recalled

While operating the machine in "Run" mode an unexcepted event may occur during the error-checking for stepper movements and pressure values. The result of this event is a false alarm being thrown and in rare cases results in an incomplete move which will affect the current inhale/exhale cycle.

Distribution information

Worldwide distribution - US Nationwide including in the state of NY and the country of Rwanda.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.