FDA recall record

Fisher & Paykel Healthcare PT301US Airvo 3 Respiratory Support Device with software version recall — FDA Z-3262-2024

Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver…

Class IIclassification
Z-3262-2024official record ID
Oct 2, 2024report date

Recall record at a glance

Official recordZ-3262-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmFisher & Paykel Healthcare, Ltd.
Report dateOct 2, 2024
Recall initiation dateAug 7, 2024
Quantity944 units
LocationAuckland, New Zealand

Official product description

PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and sub-acute facilities.

Why the product was recalled

Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.

Distribution information

U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records