FDA recall record

PryTime Medical - ER-Reboa Plus Catheter, Ref: ER7232PLUS, Guidewire Free, Compatabel with recall — FDA Z-0716-2021

There is a potential that catheters may have incorrectly marked balloon and arterial lines. This could result in a procedure delay or injury to the…

Class IIclassification
Z-0716-2021official record ID
Dec 30, 2020report date

Recall record at a glance

Official recordZ-0716-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPrytime Medical Devices, Inc.
Report dateDec 30, 2020
Recall initiation dateOct 29, 2020
Quantity3 units
LocationBoerne, TX, United States

Official product description

PryTime Medical - ER-Reboa Plus Catheter, Ref: ER7232PLUS, Guidewire Free, Compatabel with 7Fr introducer Sheaths, Arterial Line, Rx Only, UDI: (01) 0 0850001 10505 7 - Product Usage: intended for temporary occlusion of large vessels and blood pressure monitoring including patients requiring emergency control of hemorrhage.

Why the product was recalled

There is a potential that catheters may have incorrectly marked balloon and arterial lines. This could result in a procedure delay or injury to the patient.

Distribution information

US Nationwide distribution in the state of Utah.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.