FDA recall record

GE Medical Systems Information Technologies Prucka 3 Amplifiers, Model P1801PA, and Field recall — FDA Z-2613-2023

A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and…

Class IIclassification
Z-2613-2023official record ID
Sep 27, 2023report date

Recall record at a glance

Official recordZ-2613-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmGE Medical Systems Information Technologies Inc.
Report dateSep 27, 2023
Recall initiation dateAug 8, 2023
Quantity81 amplifiers
LocationWauwatosa, WI, United States

Official product description

Prucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX and ComboLab AltiX.

Why the product was recalled

A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.

Distribution information

Distribution was made to CA, CO, FL, GA, IL, IN, KY, MA, NC, NJ, OH, OR, PA, SC, TX, VA, and WI. There was government distribution but no military distribution. Foreign distribution was made to Canada, France, Germany, Greece, Ireland, Japan, Republic of Korea, and Romania.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.