Recall record at a glance
| Official record | Z-2258-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | GE Medical Systems Information Technologies Inc. |
| Report date | Jul 10, 2024 |
| Recall initiation date | Apr 22, 2024 |
| Quantity | 145 units |
| Location | Milwaukee, WI, United States |
Official product description
Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab AltiX hemodynamic and electrophysiology recording system; and Field Replacement Units spare part #5875569 (Assy CLABIII AMP 128CH 100-240V 50-60HZ).
Why the product was recalled
Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.
Distribution information
Worldwide - US Nationwide distribution in the states of AK, AZ, CA, CO, FL, GA, IL, KY, MA, ME, MN, MO, NC, NJ, NY, OH, OR, PA, TN, TX, and WA. The countries of Belgium, Canada, Czech Republic, France, Germany, Ireland, Israel, Italy, Japan, Republic of Korea, and Singapore.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.