FDA recall record

Ethicon Endo-Surgery PROXIMATE HCS Hemorrhoidal Circular Stapler Product Code: PPH01 PPH01 recall — FDA Z-0695-2019

Hemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device…

Class IIclassification
Z-0695-2019official record ID
Jan 16, 2019report date

Recall record at a glance

Official recordZ-0695-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmEthicon Endo-Surgery Inc.
Report dateJan 16, 2019
Recall initiation dateDec 7, 2018
Quantity480 units
LocationBlue Ash, OH, United States

Official product description

PROXIMATE HCS Hemorrhoidal Circular Stapler Product Code: PPH01 PPH01: The PROXIMATE HCS Hemorrhoidal Circular Stapler and Accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue.

Why the product was recalled

Hemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device may not fully cut when fired

Distribution information

Worldwide Distribution: US (Nationwide) and to countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, ECUADOR, EGYPT, INDIA, INDONESIA, JAPAN, KOREA, KUWAIT, LEBANON MEXICO, NICARAGUA, PAKISTAN, SAUDI ARABIA, SINGAPORE, TAIWAN URUGUAY and VIETNAM.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.