Leica Microsystems PROVEO 8 Ophthalmic Microscope System optical instrument for improving recall — FDA Z-0838-2019

The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or…

Class IIclassification
Z-0838-2019official record ID
Feb 27, 2019report date

Recall record at a glance

Official recordZ-0838-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmLeica Microsystems, Inc.
Report dateFeb 27, 2019
Recall initiation dateSep 25, 2017
Quantity142 units
LocationBuffalo Grove, IL, United States

Official product description

PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.

Why the product was recalled

The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or sporadic failures in the proper on-screen display functionality of the HDR recording unit. Additionally, an anomaly in the device software may also lead to the unexpected behavior that when a user releases a handle or foot switch button this may not immediately end the actuated functionality. This behavior can only occur in unlikely workflows where a user changes a user profile or a mode change is initiated at the same time as a functional button is activated.

Distribution information

Worldwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.