FDA recall record

Dain Technology ProudP Everyday Uroflow Tracker Software Version 1.x.x Software Version recall — FDA Z-1292-2022

Due to interference with the Live Listen feature of hearing aid or AirPods, the user's iPhone may perform automatic processing of the urination sound…

Class IIclassification
Z-1292-2022official record ID
Jul 6, 2022report date

Recall record at a glance

Official recordZ-1292-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmDain Technology, Inc. (Subsidiary of Soundable Health, Inc.)
Report dateJul 6, 2022
Recall initiation dateApr 21, 2022
Quantity41 devices (out of 1,260 users)
LocationSeoul, Korea (the Republic of)

Official product description

ProudP Everyday Uroflow Tracker Software Version 1.x.x Software Version 1.2.0 or earlier

Why the product was recalled

Due to interference with the Live Listen feature of hearing aid or AirPods, the user's iPhone may perform automatic processing of the urination sound signal, resulting in lower urination volume and velocity values than expected.

Distribution information

Worldwide distribution - US Nationwide. The software app is distributed via the Apple App Store.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.