Med Tec Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1 recall — FDA Z-0671-2018

When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause…

Class IIclassification
Z-0671-2018official record ID
Feb 28, 2018report date

Recall record at a glance

Official recordZ-0671-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMed Tec Inc.
Report dateFeb 28, 2018
Recall initiation dateNov 8, 2017
Quantity5 installations
LocationOrange City, IA, United States

Official product description

Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, and/or image guidance systems and the Protura Couch. The system is intended to support a treatment table and add pitch and roll to normal X, Y, Z and yaw motions to provide 6-degree axis movement. The product includes an interface between the treatment planning system and the treatment table to direct the table motion based on coordinates supplied. These products connect or support a treatment table by use of various mounting plates.

Why the product was recalled

When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to not update with the pedestal location and buffers the pedestal movement history.

Distribution information

US Distribution to Iowa and South Dakota. Shipped internationally to the Netherlands.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.