FDA recall record

Ion Beam Applications S.A ProteusPLUS and ProteusONE- designed to produce and deliver a recall — FDA Z-0747-2022

When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is…

Class IIclassification
Z-0747-2022official record ID
Mar 16, 2022report date

Recall record at a glance

Official recordZ-0747-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmIon Beam Applications S.A.
Report dateMar 16, 2022
Recall initiation dateFeb 15, 2022
Quantity15 units US; 22 units OUS
LocationLouvain-la-neuve, N/A, Belgium

Official product description

ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation Proteus 235

Why the product was recalled

When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range during a pause, there is a risk that a portion of the treatment beam after the resume is delivered with an error in range

Distribution information

IL, KS, LA, FL, MI, OK, PA, NJ, TN, TX, VA, WA Foreign: Belgium, England, Germany, France, Czech Republic, Italy, Poland, Russia, Spain, Sweden, South Korea, Taiwan, Netherlands,

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.