FDA recall record

Ion Beam Applications S.A Proteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A recall — FDA Z-2424-2018

Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in…

Class IIclassification
Z-2424-2018official record ID
Jul 18, 2018report date

Recall record at a glance

Official recordZ-2424-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmIon Beam Applications S.A.
Report dateJul 18, 2018
Recall initiation dateMay 16, 2018
Quantity3 units
LocationLouvain La Neuve, Belgium

Official product description

Proteus 235, Version PTS-11.0.1.2 and PTS-11.0.2 Product Usage: A device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Why the product was recalled

Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the corrected position.

Distribution information

Worldwide Distribution - US Nationwide in the states of FL and the countries of Sweden, The Netherlands

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.