FDA recall record

Bioporto Diagnostics A/S ProNephro AKI (NGAL) Reagent Kit (100 tests) recall — FDA Z-2937-2026

The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or…

Class IIclassification
Z-2937-2026official record ID
Aug 19, 2026report date

Recall record at a glance

Official recordZ-2937-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBioporto Diagnostics A/S
Report dateAug 19, 2026
Recall initiation dateMay 18, 2026
Quantity400 units
LocationHellerup, Denmark

Official product description

ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.

Why the product was recalled

The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users.

Distribution information

Worldwide - The affected product distribution included the United States and the countries of European Union, Chile, and Hong Kong. No affected product was distributed to Canada; however, Health Canada was notified because the product is registered in Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.