FDA recall record

Hill-Rom Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed recall — FDA Z-0326-2017

Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center…

Class IIclassification
Z-0326-2017official record ID
Nov 9, 2016report date

Recall record at a glance

Official recordZ-0326-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHill-Rom, Inc.
Report dateNov 9, 2016
Recall initiation dateAug 9, 2016
Quantity188
LocationBatesville, IN, United States

Official product description

Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.

Why the product was recalled

Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.

Distribution information

WA NC OH PA MI Australia Sweden

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceOct 6, 2021

Hill-Rom LikoStretch Mod 600 IC recall — FDA Z-2601-2021

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide…

Class IIFDA: OngoingHill-Rom, Inc.
Medical DeviceOct 6, 2021

Hill-Rom LikoStretch 1900 recall — FDA Z-2600-2021

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide…

Class IIFDA: OngoingHill-Rom, Inc.