FDA recall record

CareFusion 303 Product Name [REF]: Supply Server, CCE Enterprise SW Site License (2-5), BD recall — FDA Z-2017-2025

During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then…

Class IIclassification
Z-2017-2025official record ID
Jul 2, 2025report date

Recall record at a glance

Official recordZ-2017-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCareFusion 303, Inc.
Report dateJul 2, 2025
Recall initiation dateMay 8, 2025
Quantity6,332
LocationSan Diego, CA, United States

Official product description

Product Name [REF]: Supply Server, CCE Enterprise SW Site License (2-5), BD Pyxis Logistics [806], BD Pyxis MedStation 4000 Main 10885403512629 [303], BD Pyxis MedStation ES Tower 10885403512674 [352], BD Pyxis Anesthesia Station ES 10885403477836 [327], BD Pyxis MedStation ES (Main) 10885403512667 [323], MD BD Pyxis SupplyStation V9 [317], Security Module [139698-01], BD Knowledge Portal for Medication Technologies [136607-01], STOCKSTN V10.X WRD/WL SCANNERS [132-53-02], PRS STOCKSTN V10.X WRD/WL SCANNERS [132-53-01], STOCKSTN V10.X WIRED SCANNER [132-52-02], PRS STOCKSTN V10.X WIRED SCANNER [132-52-01], BD Pyxis CII Safe ES Desktop PC [1156-00], BD Pyxis SupplyStation Panel PC V11 [1132-00], BD Pyxis SupplyRoller V11 [1127-00], Dell 640 Server [1128-00], BD Pyxis SupplyStation System V11 [1123-00], BD Pyxis CII Safe ES 10885403512605 [1116-00], BD Pyxis Enterprise Server 10885403518348, 10885403520341 [1115-00], BD Pyxis SupplyRoller V10 [1101-00], BD Pyxis CIISafe V9.X SGL IM BIO 10885403512520 [111-221], BD Pyxis CIISafe V7.X DBL INTG MAIN BIO, SOLID DO [111-194], BD Pyxis CIISafe V8.X Desktop PC, Bio [107-245-01], BD Pyxis SupplyStation V10 [347], BD Pyxis CII Safe V9.X DESKTOP PC [107-252-01], BD Pyxis CII Safe V7.X DBL INTG MAIN BIO CLEAR DO [111-191], BD Pyxis CII Safe V7.XSGL INTG MAIN BIO SOLID DO [111-198], BD Pyxis CII SafeV8.XDBL INTG MAIN BIO SOLID DO [111-202-01], BD Pyxis CII SafeV8.XSGL INTG MAIN BIO SOLID DO [111-206-01], BD Pyxis CII Safe V9.X DBL IM BIO SLD DOORS [111-215], BD Pyxis CII Safe V9.X SGL IM BIO CLR DOORS [111-216], BD Pyxis CII Safe V9.X DBL IM BIO CLR DOORS [111-219], Supply itParActive RF V11 (Supply Kanban) [138566-01]

Why the product was recalled

During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the dispensing software application, which may lead to delay in access to medications/supplies, and delivery/replenishment of patient specific orders; which may lead to delay in patient care.

Distribution information

Worldwide - US Nationwide distribution including in the states of PR, DE, GU, MO, IA, IL, CA, MD, MA, VA, PA, NY, WA, NJ, AZ, CO, OH, FL, WI, TX, CT, LA, MI, WY, MN, AR, KY, OR, NM, TN, NC, SC, IN, VT, KS, AL, GA, WV, NV, ID, DC, MS, NE, HI, OK, UT, ND, MT, NH, SD, ME, RI, AK, MP and the countries of SA, CA, NZ , CH, FR, IT, BE, AU, BH, QA, AE, GB, EG, AR, JO, DE, BM, TH, LB, BR, IE, FI, MX, TW, ID, SG, BS, IN, MA, GF, ES, HK.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.